AERO HEAL NPWT Device

AERO HEAL NPWT Device and Kit is an advanced negative pressure wound therapy system designed for effective management of complex and high-exudate wounds. It delivers controlled and consistent negative pressure to the wound site, helping remove excess fluid, reduce bacterial load, and promote faster tissue regeneration.

This device comes with a 1000 ml canister, making it suitable for wounds with heavy drainage and for use in hospital and long-term care settings. AERO HEAL is commonly used for diabetic foot ulcers, pressure sores, post-surgical wounds, traumatic injuries, and chronic non-healing wounds. Its user-friendly design supports safe, reliable therapy with improved patient comfort and recovery outcomes.

Category:

Description

 

NPWT Device Data
NPWT Device Specifications
Model  
Device Dimensions 250×200×220mm (9.84 in × 7.87 in ×8.66 in) (without canister)
Material ABS
Weight 2.8kg (~6.16 lbs.) (without canister and adaptor)
Canister bag Volume 1000ml
Dimensions 195×85×85mm (7.68 in×3.35 in×3.35 in)
Material  plastic
Weight 185g  (~0.41 lbs)
Used capacity ≥1000ml×90%
Display screen size 5.5 inch
Display  Colorful Touch Screen
Drainage tube length ≥1750mm
Normal working noise ≤55dB
Pressure options
Pressure of Machine -40~-300mmHg(-5.3~-39.9kPa)
Pressure of Canister -225mmHg±10mmHg(-29.9kPa±1.3kPa)
Anti-overflow During normal use, the exudate in the drainage bottle cannot be sucked into the negative pressure machine
Limit negative pressure Not less than -330mmHg(-44kPa)
Gas flow <8000ml/min
Therapy Delivery Modes Continuous Mode/Intermittent Mode/Dynamic Mode
Alarm function Air leakage alarm/Blockage pipes alarm/Low charge alarm
Electrical Parameters
External Power Supply Input AC100~240V,50/60Hz
External Power Supply Output DC18V ,≥2A
Power Adapter cable length ≥1.2m
Battery  Parameters
External Power Supply Input 100~240V,50/60Hz
External Power Supply Output 11.1V (3.7V ×3), 3.2A
Battery Run Life ≥ 8hours
Battery Charge Time 3.5 hours from a fully discharged state or it takes about 2 hours to charge the battery to 80% of the total battery capacity.
Environmental Conditions
Storage Conditions
Temperature Range -10℃~50℃
Relative Humidity Range ≤95% R.H
Others No corrosive gas, good ventilation
Operating Conditions
Temperature Range 5℃~40℃
Relative Humidity Range ≤80% R.H
Atmospheric pressure 50kPa~110kPa
Expected Service Life and Warranty Period
Expected Service Life 3 years
Warranty period 2 years

 